Pacemaker
|
|
| Pinnacle I ( Infinite) Long Life Pacemaker | | Programmable Parameters | | Pacing Modes | VVI, VVT, VOO, AAI, AAT, AOO, OVO,OAO | | Basic Pacing Rates | From 30 to 120 Min -1 | | Pulse Widths | From 0.07 ms to 1.5 ms | | Pulse Amplitudes | 0.2 V to 5.0 V in 32 steps | | Refractory Periods | From 203 to 438 ms 16 values/td> | | Sensitivities | From 0.5 to 4.0 mV | Hysterisis | From 0 to 20% OF RATE | Trigger Upper Rate | 78 -183 Min-1 | | | | | | Other Parameters | | | Magnet Rate | 96 Min -1 at BOL, 86 Min -1 at ERI | | Battery life at 100% pacing | 60 Min -1, 3.6V, 0.4 ms, 3.0 mV | | At BOL | 13.5 uA, 15.2 years | | At ERI | 8.5 uA | | Battery life at 100% inhibited/td> | 60 Min -1, 3.6V, 0.4 ms, 3.0 mV | | At BOL | 10.0 uA, 20 years | | At ERI | 8 uA | | Battery life at 100% pacing | 60 Min -1, 2.5V, 0.4 ms, 3.0 mV | | At BOL | 11uA, 19 years | | At ERI | 8.0 uA | | Battery Chemistry | LI 3178 / WG 8206 Implantable Li I 1.9 Ah (1.88 usable) | | Polarity | Bipolar, Unipolar | | Lead | Universal IS-1 (3.2 mm) 7F | | | | |
| |
Programmable Parameters | | Real Time | Yes | | Battery Status | Yes | | Marker Pulses | Yes | | Serial Number | Yes | | Impedance Measure | Yes | | Pacing Threshold | Yes | Sensing Threshold | Yes | Temporary Programming | Yes | Emergency Programming | Yes | Pacing & Sensing Counts | Yes | | |
Physical Characteristics | | Width | 37 mm x 43 mm x 9.3 mm | | Mass | 30 g & 11.5 cc | | |
|
|