Tiranga 302 BI/ 301 Uni 3 Parameter Programmable Pacemaker
Specifications:
General parameter:
- Pacing Modes: VVI, VVT, VOO, AAI, AAT, AOO, VVIR, VOOR, VVTR, AAIR, AOOR, AATR, OAO, OVO, OFF
- Basic(Lower) Rate: From 30 BPM to 120 BPM
- Upper Sensor Rate: From 80 BPM to 180 BPM in steps of 2 BPM.
- Upper Rate: From 80 BPM to 180 BPM in steps of 2 BPM.(In SST Mode)
- High Rate Temporary programming
- Auto Measure (Auto threshold Measurement) (ON/OFF)
Pacing and sensing Parameters:
- Pulse Width: From 0.07 ms to 1.5 ms.
- Amplitude: From 0.2 V to 7.5 V.
- Pacing Polarity: Unipolar /Bipolar
- Sensitivity: 0.5 to 7.5 mV
- Sensing Polarity: Unipolar/Bipolar
- Refractory Period: From 200 ms to 500 ms in steps of 25 ms
Hysteresis Parameter
- Hysteresis Enable : On/Off
- Hysteresis Rate Change : From 2 BPM to 40 BPM in steps of 2 BPM
- Hysteresis Search: On/Off.
Activity Parameters:
- Upper Activity Rate: From 80 BPM to 180 BPM in steps of 2 BPM.
- Activity Rate Response: From 1 to 15
- Reaction Time: From 10 s to 60 s in steps of 10 s
- Recovery Time: From 1 m to 10 m in steps of 1 m.
- Automatic Refractory Change: On/Off. When programmed On, if the rate is greater than 120 BPM, the refractory switch to the minimum value between the programmed value and 250 ms.
Magnetic Response:
- BOL: SOO mode. Rate 96 BPM.
- ERI: SOO mode. Rate 84 BPM.
- EOL: SOO mode. Rate 64 BPM
Battery Chemistry and Model: Lithium Iodine WG9085 (0.98Ah).
Initial Voltage: 2.8V.
Service Life: The expected life of the Pacemaker, pacing 100% at 60 min-1, SSI mode, with 2.5 V
amplitude, pulse width of 0.4 ms and lead impedance 500 Ohms, is more than 7 Years.
Size: 44X38X11 mm
Mass: 23 gm
|
Physical Characteristics |
|
Dimensions |
44 x 38 x 7.4 mm |
Mass |
.26 g |
Case Material |
Titanium |
Connectors |
3.2mm, IS-1 |
|
|
Service Life
The expected life of the generator, pacing 100% at 70 ppm, with 3.6 V amplitude and pulse width of 0.4 ms, is 9.2 years.
The expected life of the generator, pacing 100% at 60 ppm, with 2.5 V amplitude and pulse width of 0.4 ms, is 12.6 years.
Telemetry Data
Parameters: All the programmable parameters.
Identification: Serial number
Battery condition: Voltage, ERI and EOL indicators Threshold test by Pulse width and sensing available.
Statistics: Keeps track of pacing and sensing events, as well as ventricular premature events.
Marker Pulses.
Protections
Noise detection: When the pacemaker senses events with frequency exceeding 11 Hz, it switches to VOO mode.
Run Away Limit: The pacemaker includes an antirunaway circuit to avoid the generation of stimuli with a rate over 200 ppm (except in temporary programming).
Programming
The pacemaker is programmed using a Laptop that runs Windows 98 / 2000 / XP operating system. The programming system includes Log of Pacemaker History. It also permits the measurement of the stimulation thresholds using Pulse width. Top